Terumo Capiox NX19 Oxygenator (East Orientation), P/N: CXOT280 Exported only to Japan for inclusion into convenience kits that are sold only within the Japanese market
FDA Recall Z-2103-2019 · Posted None
Advertisement
Recalling FirmTerumo Cardiovascular Systems Corporation
Recall StatusTerminated
Product CodeDTZ
Date Initiated2019-06-14
Date Terminated2020-06-11
Product Quantity78 units
Reason for Recall
Capiox NX19 Oxygenator may contain a non-functioning arterial thermistor component
Firm Action
Terumo issued Urgent Medical Device Notification letter dated June 13 2019 via FedEx mailing executed on 6/14/2019 advising of the problem, health risk and action to take: Terumo CVS advises users to remove all affected product from inventory and return unused product to Terumo CVS following the instructions provided in this notice.
Questions or concerns:
Terumo CVS Customer Service: 1.800.521.2818 Monday Friday, 8 a.m. 6 p.m. ET
Terumo Recall Fax: 1.410.392.7183
Terumo Recall Email: [email protected]
Root Cause
Process control
Distribution Pattern
MI Foreign: Japan, Belgium
Related FDA 510(k) Records