Open, Classified

VersaCross Access Solution (VXSK), Product Description and ID: VXSK (180J RF WIRE 63CM/45DEG/D0) VXSK0101 VXSK (180J RF WIRE 63CM/45DEG/D1) VXSK0102 VXSK (180J RF WIRE 63CM/55DEG/D0) VXSK0103 VXSK (180P RF WIRE 63CM/45DEG/D0) VXSK0121 VXSK (180P RF WIRE 63CM/45DEG/D1) VXSK0122 VXSK (180P RF WIRE 63CM/55DEG/D0) VXSK0123 VXSK (180P RF WIRE 63CM/55DEG/D1) VXSK0124 VXSK (180P RF WIRE 63CM/90DEG/D0) VXSK0125 VXSK (230J RF WIRE 63CM/45DEG/D0) VXSK0111 VXSK (230J RF WIRE 63CM/45DEG/D1) VXSK0112 VXSK (230P RF WIRE 63CM/45DEG/D0) VXSK0131 VXSK (230P RF WIRE 63CM/45DEG/D1) VXSK0132 VXSK (230P RF WIRE 63CM/55DEG/D0) VXSK0133 VXSK (230P RF WIRE 63CM/55DEG/D1) VXSK0134 VXSK (230P RF WIRE 81CM/45DEG/D0) VXSK0137 VXSK-180P RF WIRE 63CM/45DEG/D0 SHEATH VXSK0021 VXSK-180P RF WIRE 63CM/45DEG/D1 SHEATH VXSK0022 VXSK-180P RF WIRE 63CM/55DEG/D1 SHEATH VXSK0024 VXSK-180P RF WIRE 63CM/90DEG/D0 SHEATH VXSK0025 VXSK-230P RF WIRE 63CM/45DEG/D0 SHEATH VXSK0031 VXSK-230P RF WIRE 63CM/45DEG/D1 SHEATH VXSK0032 VXSK-230P RF WIRE 63CM/55DEG/D1 SHEATH VXSK0034

FDA Recall Z-2103-2025 · Posted 2025-07-10

Recall StatusOpen, Classified
Product CodeDXF
Date Initiated2025-06-10
Product Quantity14,724

Reason for Recall

Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.

Firm Action

On June 10, 2025, the firm began notifying consignees of the field action via Urgent Medical Device Removal letters. Customers are instructed to not use affected product. They should remove affected units from their inventory and segregate the units in a secure place until they can be returned to Boston Scientific.

Root Cause

Packaging change control

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Canada, Hong Kong, and Japan.