Boston Scientific Flexima Biliary Stent System, 200 cm long, .035" (0.89 mm), sterile, Boston Scientific, Spencer, IN; REF 3921, UPN M00539210. System is used for delivery of the stent to the biliary tract for drainage of the bile ducts, for splinting of a bile duct during healing, or for providing bile duct patency in a stricture or past a stone.

FDA Recall Z-2105-2008 · Posted 2008-09-11

Recall StatusTerminated
Product CodeFGE
Date Initiated2008-04-02
Date Terminated2008-09-11
Product Quantity80

Reason for Recall

Rapid Exchange Biliary Stent may be mislabeled as Flexima biliary Stent, which has a different delivery system included in the package.

Firm Action

An Urgent Medical Device Recall-Immediate Action Required letter was sent to consignees on 4/2/08 directing them to immediately discontinue use of and segregate the identified products for return to the firm and complete the verification tracking form.

Root Cause

Packaging process control

Distribution Pattern

Class III Recall - Worldwide Distribution --- including USA and countries of Finland, France, Germany and Great Britain.