14 mm Distraction Pin, Exagg Qty: 1 per pack; STERILE; Manufactured for: Medline Industries, Inc., Mundelein, IL 60060 USA Distraction Pin is designed for Anterior Cervical Fusion Procedures. Use with vertebral body distraction instruments. It is intended as a temporary fixation screw utilized during cervical spine procedures.

FDA Recall Z-2106-2017 · Posted None

Recall StatusTerminated
Product CodeLXH
Date Initiated2016-12-14
Date Terminated2019-07-09
Product Quantity1031 individual packs

Reason for Recall

Product's non-conformity involves the integrity of the seal in the sterile packaging. It is possible that the seal on the sterile packaging has been compromised resulting in a loss of sterility of the medical device contained within.

Firm Action

Medline Industries sent an Immediate Action Required letter dated December 14, 2016, to all affected customers with response forms via US mail, notifying them of the recall. Customers were instructed to quarantine affected product and return it to the firm. The product will be repackaged and sterilized. Customers with questions were instructed to call 866-359-1704. For questions regarding this recall call 847-643-3245.

Root Cause

Process control

Distribution Pattern

Nationwide Distribution