ARTIS Pheno - Interventional Fluoroscopic X-Ray System Model: 10849000 - Product Usage: is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

FDA Recall Z-2108-2020 · Posted None

Recall StatusTerminated
Product CodeOWB
Date Initiated2020-04-22
Date Terminated2020-10-09
Product Quantity105 systems

Reason for Recall

If the C-arm leaves its intended travel path due to a fault within the drivetrain, movements may be impacted or impossible and the area of interest cannot be reached. In the event the C-arms movements are not possible, the system can only be returned to normal operation with the support of a field service engineer

Firm Action

Siemens issued Urgent Medical Device Correction dated 4/22/20 via AX074/19/S. Letter states reason for recall, health risk and action to take: Planned procedures may have to be terminated and performed on an alternative X-ray system. Please follow the standard emergency procedures you have in place in the event of system failures. It is recommended to consider reinforcing these emergency procedures until the update has been completed. Siemens will correct the error with a software patch in which the system will recognize a deviation of the intended travel path and unit movements are stopped immediately, similar to the activation of a proximity switch. The correction will be initiated via Update Instruction AX073/19/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the correction, the cause is eliminated, and any recurrence of this potential fault is prevented

Root Cause

Software design

Distribution Pattern

US Nationwide distribution.