Terminated

Minor Surgical Kits, Sterile-Kits containing the Devon Light Glove Catalog Numbers/Description: 573208 7519-DHCG MINOR KIT 573210 7581-CSD Minor KIT 573741 7667-HMI Minor Kit 31144507 7519 Minor Surgical Kit 573830D MINOR KIT

FDA Recall Z-2109-2015 · Posted 2015-07-20

Recalling FirmCovidien LLC
Recall StatusTerminated
Product CodeFSY
Date Initiated2015-04-16
Date Terminated2017-06-28
Product Quantity81,252 kits

Reason for Recall

Devon Light Gloves contain splits or holes compromising the sterility

Firm Action

Medtronic/Covidien initiated recall on April 16, 2015 via FedEx and certified mail Customers were instructed to discontinue use of the product and return all inventory and kits to Medtronic (Covidien) -Attention: Devon" Light Glove 110 Kendall Park Lane. Atlanta, GA 30336. If you have any questions contact your Medtronic representative or Medtronic Customer Service, Monday through Friday, 8am to 6:30pm ET, at (800) 882-5878.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution-US (nationwide) including the Virgin Islands and the countries of Canada, Australia, New Zealand, Costa Rica, Austria, Belgium, Bulgaria, Czech Republic, Denmark, Finland , France, Germany, Netherlands, Sweden, Ireland, ISRAEL,Spain, Italy, Jordan, Kenya, Kuwait, United Kingdom, Lebanon, Liechtenstein, Luxembourg, Monaco, South Africa, Norway, Poland, Portugal, Qatar, Russia, Saudi Arabia, Slovenia, Switzerland, Turkey, and UAE.