Terminated

Big Bore Oncology tomography X-ray system scanner, model number 728243. The Big Bore Oncology scanner is a whole body computed tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes

FDA Recall Z-2109-2016 · Posted None

Recall StatusTerminated
Product CodeJAK
Date Initiated2016-06-14
Date Terminated2018-07-02
Product Quantity47 units

Reason for Recall

Socket Head Cap Screws (SHCS) used in the Z-Axis Adjustment Plate were found to be made of stainless steel rather than the specified alloy steel.

Firm Action

Philips Medical Systems (Cleveland), Inc. sent a Customer Information letter dated June 14, 2016, to all affected customers via certified mail. The letter identified the product, the problem, and the action to be taken by the customer. No action is required of the customer for this issue. Systems that may have been manufactured with these stainless steel fasteners still meet the safety specifications required for the adjustment cradle and may continue to be used clinically. For further information or support customers were instructed to call 1-800-722-9377. For questions regarding this recall call 404-483-2015.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide distribution to Belgium, China, Egypt, Germany, India, Korea, Republic of, Myanmar, Nigeria, Peru, Russian Federation, Rwanda, Singapore, Spain, Sweden, Taiwan, Thailand, and United Kingdom