Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Surgical Set Up kit, Sterile-Kits containing the Devon Light Glove Catalog Numbers/Description: 31144895 7614-T4 Single Basin Set Up Kit 31144960 7693-T4 Single Basin Set Up Kit 31144978 7766 Double Basin Set Up Kit 31145215 7896 Double Basin Set Up Kit 31145231 7897 Double Basin Set Up Kit 31145249 7897-T8 Double Basin Set Up Kit 31145298 7896-T8 Double Basin Set Up Kit 31145496 7696-T4 Single Basin Set Up Kit 31145520 7614 Single Basin Set Up Kit 31145546 7693 Single Basin Set Up Kit 31145629 7696 Single Basin Set Up Kit 31153938 7756-KST Surgical Set Up Kit 31154266 7656-KST Surgical Set Up Kit 31321097 7697 Surgical Set Up Kit 31324299 7830-HOH Surgical Set Up Kit 31451092 7600-DNV Surgical Set Up Kit 31453098 7682-MHP Surgical Set Up Kit

FDA Recall Z-2110-2015 · Posted 2015-07-20

Recalling FirmCovidien LLC
Recall StatusTerminated
Product CodeFSY
Date Initiated2015-04-16
Date Terminated2017-06-28
Product Quantity33,256 kits

Reason for Recall

Devon Light Gloves contain splits or holes compromising the sterility

Firm Action

Medtronic/Covidien initiated recall on April 16, 2015 via FedEx and certified mail Customers were instructed to discontinue use of the product and return all inventory and kits to Medtronic (Covidien) -Attention: Devon" Light Glove 110 Kendall Park Lane. Atlanta, GA 30336. If you have any questions contact your Medtronic representative or Medtronic Customer Service, Monday through Friday, 8am to 6:30pm ET, at (800) 882-5878.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution-US (nationwide) including the Virgin Islands and the countries of Canada, Australia, New Zealand, Costa Rica, Austria, Belgium, Bulgaria, Czech Republic, Denmark, Finland , France, Germany, Netherlands, Sweden, Ireland, ISRAEL,Spain, Italy, Jordan, Kenya, Kuwait, United Kingdom, Lebanon, Liechtenstein, Luxembourg, Monaco, South Africa, Norway, Poland, Portugal, Qatar, Russia, Saudi Arabia, Slovenia, Switzerland, Turkey, and UAE.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑