Terminated

Version: 1.2.0 (1.2.0+85ela410) Caption Health, Caption AI, Ultrasound Imaging System Application Software (GMDN: 40873), UDI: (01)00860003586303(10)1.2.0)

FDA Recall Z-2111-2021 · Posted None

Recalling FirmCaption Health
Recall StatusTerminated
Product CodeQIH
Date Initiated2021-05-19
Date Terminated2023-01-31
Product Quantity15 systems

Reason for Recall

The firm is aware of a issue with ultrasound software that results in unintended video frames being included. This could lead to an incorrect automated ejection fraction (AutoEF) and image quality score (IQS) estimates.

Firm Action

On May 19, 2021 Caption health, Inc. contacted consignees via telephone to inform them about their recall. On May 20, 2021, the firm sent follow-up communication via via email. In addition to informing consignees about the recall, the firm also asked customers to take the following actions: 1. Until your system is updated, we recommend that all EF calculations are reviewed carefully using clinical expertise, to verify the accuracy of the results. 2. If you have any questions or concerns, please contact Caption health support line at 1-855-506-HELP between 7am-7pm ET. You can also email us at [email protected] at any time. We will follow-up with additional details relevant to your site by Monday May 24

Root Cause

Software design

Distribution Pattern

US : AZ, DE, NY, IL, CT OUS: None