Padded Swing Away Arm Rest with Receiver; Uses on wheelchair models: Quickie Q7 (EIR4), GT (EIR3), GTi (EIT4) and Zippie Zone (EIRZ1). Product Usage: To empower physically challenged persons by providing a means of mobility. This includes temporary and permanent conditions.

FDA Recall Z-2113-2012 · Posted 2012-07-31

Recall StatusTerminated
Product CodeIOR
Date Initiated2012-06-26
Date Terminated2012-11-01
Product Quantity14, 035

Reason for Recall

Reports of injury when the Padded Swing Away Armrest is used for full body weight. Owner's manual states that armrest is only designed for resting of forearm.

Firm Action

Sunrise Medical sent a Safety Notice letter dated June 25, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter request that clinicians/suppliers post this safety notice in a conspicuous location, making customers and therapists aware of the issue. The letter also encourage end users to read the owner's manual and follow the recommenations for proper usage of the recalled product. Customers are instructed to complete and return the Supplier Notification and Response Form. For questions contact Sunrise Medical Regulatory Affairs at 888-208-4901.

Root Cause

Labeling design

Distribution Pattern

Worldwide Distribution - US (nationwide) and the countries of Argentina, Brazil, Columbia, Costa Rica, Mexico, Panama, Uruguay and Venezuela.