Ingenuity Core, model # 728326; Intended to produce cross- sectional images of the body.

FDA Recall Z-2115-2016 · Posted None

Recall StatusTerminated
Product CodeJAK
Date Initiated2016-06-21
Date Terminated2018-07-02
Product Quantity1 unit

Reason for Recall

Supplier design change contributed to a thinner plastic which was used as part of the cover on the devices. The thinner plastic cover may crack causing injury to patients, operators, by-standers and service personnel.

Firm Action

Philips is informing the consignee of this issue through: Customer Information Letter (CIL) 72800656 . International consignees will be contacted by Philips Healthcare Global Market representatives, who will distribute the CIL. A Field Service Engineer (FSE) will contact the customers to schedule a time to conduct the appropriate correction through Field Change Order (FCO) 72800656 once it is available. This FCO will address the above identified issue. A response form sent to the customer via certified mail. This package will also contain a return receipt.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Distributed in China, Denmark, Germany, Hungary, Italy, Philippines, Spain, Sweden, Switzerland, Thailand, and United Kingdom.