Terminated

EG-3630U Ultrasound Video Gastroscope EG-3830UT Ultrasound Video Gastroscope EG-3870UTK Ultrasound Video Gastroscope FG-36UX Ultrasound Fiber Gastroscope Used to provide optical and sonographic visualization of and therapeutic access to the Upper Gastrointestinal Tract.

FDA Recall Z-2116-2014 · Posted 2014-07-24

Recall StatusTerminated
Product CodeFDS
Date Initiated2014-06-12
Date Terminated2015-11-17
Product Quantity531 units

Reason for Recall

Aspiration needles used in combination with Ultrasound Gastroscopes can derail from the accessory elevator at the distal end of the endoscope. This may result in patient injury if the Instruction for Use is not followed carefully by the user.

Firm Action

Pentax Medical sent an Urgent Medical Device Field Safety Corrective Action dated June 12, 2014 to affected customers. The letter identified the affected product, problem and actions to be taken. Pentax also posted their notification on their website. For questions regarding this action please feel free to contact Paul Silva, Recall Coordinator at Tel: 800-431-5880 ext 2064 (8:30am 5:00pm Mon-Fri, EST) Fax: 201-799-4063 Email : [email protected]

Root Cause

Device Design

Distribution Pattern

US Nationwide Distribution