Terminated

VACUETTE Blood Transfer Unit. Lot no. A19024NX. Expiration date: 2022-02-07 - Product Usage: The Blood Transfer Unit is used to transfer blood from a syringe into the evacuated blood collection tube.

FDA Recall Z-2116-2020 · Posted None

Recall StatusTerminated
Product CodeJKA
Date Initiated2020-04-03
Date Terminated2021-07-25
Product Quantity90,400 pieces.

Reason for Recall

While transferring blood the needle may dislodged and remain in the tube stopper which may cause needle stick injury or blood leakage.

Firm Action

Greiner BIO-ONE notified customers on 04/03/2020 via "Urgent Product Recall-VACUETTE Blood Transfer Unit" letter. The recall letter identified the affected product and lot numbers. The firm requested the customers to immediately isolate the defective product and complete the product disposition form.

Root Cause

Under Investigation by firm

Distribution Pattern

*requested 4/15/2020

Official Source

View FDA Recall →