Siemens Healthcare Diagnostics, Dimension Vista LOCI Reaction Vessels, catalog number KS855. For in vitro diagnostic use.

FDA Recall Z-2117-2010 · Posted 2010-08-02

Recall StatusTerminated
Product CodeJJE
Date Initiated2010-03-15
Date Terminated2010-08-03
Product Quantity5544 bags

Reason for Recall

Vessel pick up errors and vessel jams when the vessel is being loaded. The error halts the instrument and disrupts processing. Vessel flange is thin.

Firm Action

Siemens Healthcare Diagnostics issued an Urgent Field Safety Notice dated March 2010 informing users of the potential problem, to discard any remaining product, and if needed, to request a replacement lot.

Root Cause

Process design

Distribution Pattern

Worldwide distribution: USA, Canada, Europe, Japan, Korea, and New Zealand.