Locking Screw, D 2.0MM X 8MM, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label. Size (Part Number): a) D 2.0MM X 8MM (002-E-20008); b) D 2.0MM X 10MM (002-E-20010); c) D 2.0MM X 12MM (002-E-20012); d) D 2.0MM X 14MM (002-E-20014); e) D 2.0MM X 16MM (002-E-20016); f) D 2.0MM X 18MM (002-E-20018); g) D 2.0MM X 20MM (002-E-20020); h) D 2.0MM X 22MM (002-E-20022); i) D 2.0MM X 24MM (002-E-20024); j) D 2.0MM X 26MM (002-E-20026); k) D 2.0MM X 28MM (002-E-20028); l) D 2.0MM X 30MM (002-E-20030)
FDA Recall Z-2120-2018 · Posted 2018-06-07
Advertisement
Recalling FirmOrthofix, Inc
Recall StatusTerminated
Product CodeHRS
Date Initiated2018-04-06
Date Terminated2021-06-30
Product Quantity870
Reason for Recall
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Firm Action
On 04/06/2018, Orthofix Inc. (U.S. Agent), sent written notification of the recall to distributors and hospitals via e-mail and/or FedEx overnight courier.
Root Cause
Device Design
Distribution Pattern
U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,