Terminated

Non-Locking Screw, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a)D 3.2MM X 10MM (002-D-32010) b) D 3.2MM X 12MM (002-D-32012) c) D 3.2MM X 14MM (002-D-32014) d) D 3.2MM X 16MM (002-D-32016) e) D 3.2MM X 18MM (002-D-32018) f) D 3.2MM X 20MM (002-D-32020) g) D 3.2MM X 22MM (002-D-32022) h) D 3.2MM X 24MM (002-D-32024) i) D 3.2MM X 26MM (002-D-32026) j) D 3.2MM X 28MM (002-D-32028) k) D 3.2MM X 30MM (002-D-32030) l) D 3.2MM X 32MM (002-D-32032) m) D 3.2MM X 34MM (002-D-32034) n) D 3.2MM X 36MM (002-D-32036) o) D 3.2MM X 38MM (002-D-32038) p) D 3.2MM X 40MM (002-D-32040) q) D 3.2MM X 42MM (002-D-32042) r) D 3.2MM X 44MM (002-D-32044) s) D 3.2MM X 46MM (002-D-32046) t) D 3.2MM X 48MM (002-D-32048) u) D 3.2MM X 50MM (002-D-32050)

FDA Recall Z-2123-2018 · Posted 2018-06-07

Recalling FirmOrthofix, Inc
Recall StatusTerminated
Product CodeHRS
Date Initiated2018-04-06
Date Terminated2021-06-30
Product Quantity1206

Reason for Recall

Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.

Firm Action

On 04/06/2018, Orthofix Inc. (U.S. Agent), sent written notification of the recall to distributors and hospitals via e-mail and/or FedEx overnight courier.

Root Cause

Device Design

Distribution Pattern

U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,

Official Source

View FDA Recall →