BARD DORADO PTA BALLOON DILATATION CATHETERS, PRODUCT CODE # DR135417
FDA Recall Z-2124-2010 · Posted 2010-08-02
Advertisement
Recalling FirmBard Peripheral Vascular Inc
Recall StatusTerminated
Product CodeDQY
Date Initiated2010-05-04
Date Terminated2012-02-01
Product Quantity213
Reason for Recall
The affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft.
Firm Action
The Bard Peripheral Vascular "URGENT MEDICAL DEVICE RECALL NOTIFICATION" with attached "RECALL AND EFFECTIVENESS CHECK FORM" was sent to the affected consignees beginning May 7, 2010 via FedEx. The recall notice informed the consignees of the reason for recall, the products affected and instructed them not to use or further distribute any affected product.
The consignees were also instructed to call the firm's Recall Coordinator Silvia Carrillo at 1-800-321-4254 Option #2 Ext 2727 (M-F 7am to 4pm MST) or email at [email protected] to facilitate the expedient return of the affected products. Additionally they were asked to fax the Recall and Effectiveness Check Form to BPV at 1-800-994-6772.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Nationwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.