KWIK-STIK and LYFO DISK, quality control kit for culture media Catalog # 0805K and 0805L

FDA Recall Z-2124-2025 · Posted 2025-07-17

Recalling FirmMicrobiologics Inc
Recall StatusOpen, Classified
Product CodeJTR
Date Initiated2025-06-17
Product Quantity21 total

Reason for Recall

Affected products contain Staphylococcus epidermidis instead of the expected Brevundimonas diminuta. This may cause the user's quality control to fail and delayed diagnosis for the patient.

Firm Action

On June 17. 2025, the firm began notifying affected customers via "Urgent Medical Device Recall" letters delivered through email. Customers were instructed to review lab procedures to understand how the information may affect usage; use or discard product depending on lab procedures. Complete and return the response form to [email protected]. If product was further distributed or transferred to other facilities or customers, they should also be notified. Customers may contact Microbiologics if a replacement kit is needed. If you have any questions or concerns, contact Recall Support team at 320.229.7080 or [email protected]. Collect calls may be made.

Root Cause

Under Investigation by firm

Distribution Pattern

US distribution to states of: Utah and Ohio; and OUS (International) to countries of: Italy, Netherlands, Argentina, Korea, Jordan, Spain, and Bangladesh.