Ceftriaxone TX 32 US S30, Ref 412302 and Ceftriaxone TX 32 US F100, Ref 507058 ETEST is a quantitative technique for determining the antimicrobial susceptibility of Gram-negative and Gram-positive aerobic ba

FDA Recall Z-2126-2017 · Posted 2017-05-03

Recalling FirmBioMerieux SA
Recall StatusTerminated
Product CodeJWY
Date Initiated2017-01-01
Date Terminated2022-02-24
Product Quantity4656 kits

Reason for Recall

Potential performance issue on strain categorization.

Firm Action

Biomerieux sent an Urgent Field Safety Notice dated January 2017, to Laboratory Manager or Laboratory Director. The Notice indicated what actions to take and included a reply form to be returned. Required actions: The following recommendations require your immediate attention to ensure the product will continue to perform per its labeled performance specifications, within its revised shelf-life of twelve (12) months. Product with NO remaining shelf life (after reduction): Identify impacted lots of ETEST Ceftriaxone TXL32 (Ref. 412302, 412303) (lots listed in Table 1 below) which are now expired after shelf-life reduction. Immediately order the replacement products appropriate for your institution. Until replacement product is available Laboratories may continue to use their now expired strips with the following recommendations: Laboratories should continue to follow their current QC procedures for ETEST Ceftriaxone TXL32 (Ref. 412302, 412303) for the lots listed in Table 1 in accordance with CLIA and local regulatory requirements, with a modification to increase the frequency of QC testing to weekly or every day of use if previous QC testing exceeds one week. Laboratories should include in the QC testing the strain N. gonorrhoeae ATCC 49226 (Expected range-MIC: 0.004-0.016 g/mL) defined as the stability indicator for ETEST Ceftriaxone TXL32 (Ref. 412302, 412303). The MIC result for this specific strain must fall in the acceptable range to confirm the validity of the QC test and performance of the ETEST with the clinical isolate. Only report results if all QC is in the acceptable ranges. When replacement product is received, discontinue using and discard impacted ETEST Ceftriaxone TXL32 (Ref. 412302, 412303) (listed in Table 1). Additional actions: Please distribute this information to all appropriate personnel in your laboratory, retain a copy in your files, and forward this information to all parties that may use

Root Cause

Other

Distribution Pattern

USA (nationwide) Distribution to the states of : AZ, AR, CA, CT, DE, FL, GA, HI, IL, LA, ME, MA, MI, MN, MO, MT, NH, NY, NC, OR, PA, SC, SD, TN, TX, WA