GE Healthcare Vivid S60N, ultrasound device

FDA Recall Z-2126-2023 · Posted 2023-07-10

Recall StatusOpen, Classified
Product CodeIYN
Date Initiated2023-05-30

Reason for Recall

GE HealthCare has become aware that certain Vivid ultrasound systems cannot boot up in a timely fashion. If this occurs, it can delay availability of the device in time-critical emergency situations.

Firm Action

GE Healthcare sent an Urgent Medical Device Correction notice to its consignees on 05/30/2023 by mail (traceable means). The notice explained the problem with the instrument and provided the following: "You can continue to use your device. Please follow clinical practice guidelines, which include having a backup imaging plan when performing time-critical examinations or image-guided interventions." The devices will be corrected at no charge to the consignee.

Root Cause

Process design

Distribution Pattern

Worldwide distribution.