Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Brand Name: EMPOWR 3D KNEE TIBIAL INSERT - E-PLUS Product Name: EMPOWR 3D KNEETM INS, 11L 12MM, VE Model/Catalog Number: 341-12-711 Software Version: NA Product Description: Enovis" knee devices are intended for treatment of patients who are candidates for knee arthroplasty per the Indication for use. While total knee replacements are not intended to withstand activity levels and loads of normal healthy bone, they are a means of restoring mobility and reducing pain for many patients. Component: NA

FDA Recall Z-2126-2025 · Posted 2025-07-17

Recalling FirmEncore Medical, LP
Recall StatusOpen, Classified
Product CodeJWH
Date Initiated2025-06-02
Product Quantity20

Reason for Recall

Due to packaging discrepancy with the incorrect tibial inserts/implants being packaged. The use of this product may result in the incorrect product being selected. This could result in a revision surgery if used.

Firm Action

On 06/02/2025, the firm emailed an "Urgent: Medical Device Recall" letter to customers. The letter described the product, problem and actions to be taken. Customers are instructed to: 1.Review their stock for the products and lot numbers for the items to be returned. 2.This notice needs to be passed on to all those within their organization or to any organization where the product has been transferred. 3.Complete acknowledgement and response form and contact Customer Service at 1-800-456-8696 or through their regional email address, to place a replacement request. They will be provided with an RMA number and shipping label to return the affected product. 4.Return all devices (quantities listed above) using the RMA number and shipping label provided by Customer Service. Any questions, contact Recall Coordinator at [email protected] or call/text 682-299-9133.

Root Cause

Packaging change control

Distribution Pattern

U.S Distribution to states of: AZ, IA, ID, IL, IN, LA, MI, MO, MS, NE, NY, OK, OR, PR, SD, UT, VA, and WA

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑