BARD DORADO PTA BALLOON DILATATION CATHETERS, PRODUCT CODE # DR135520

FDA Recall Z-2129-2010 · Posted 2010-08-02

Recall StatusTerminated
Product CodeDQY
Date Initiated2010-05-04
Date Terminated2012-02-01
Product Quantity522 units

Reason for Recall

The affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft.

Firm Action

The Bard Peripheral Vascular "URGENT MEDICAL DEVICE RECALL NOTIFICATION" with attached "RECALL AND EFFECTIVENESS CHECK FORM" was sent to the affected consignees beginning May 7, 2010 via FedEx. The recall notice informed the consignees of the reason for recall, the products affected and instructed them not to use or further distribute any affected product. The consignees were also instructed to call the firm's Recall Coordinator Silvia Carrillo at 1-800-321-4254 Option #2 Ext 2727 (M-F 7am to 4pm MST) or email at [email protected] to facilitate the expedient return of the affected products. Additionally they were asked to fax the Recall and Effectiveness Check Form to BPV at 1-800-994-6772.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Nationwide