da Vinci SP Camera; PN 430060-33; Model Number: 430060-33. For endoscopic surgery.
FDA Recall Z-2131-2020 · Posted None
Advertisement
Recalling FirmIntuitive Surgical, Inc.
Recall StatusOpen, Classified
Product CodeNAY
Date Initiated2019-08-19
Product Quantity305 units
Reason for Recall
Some cameras were found to have residual soil adhered to the tip of the camera after reprocessing which may affect image quality resulting in needing to switch to a second camera or converting to another surgical modality. If residual soil makes contact with the patient, exposure could result in sensitization making future component blood therapy or tissue typing more challenging; additionally if a small piece of inert epoxy were to be retained, scar tissue could form.
Firm Action
Urgent Medical Device Recall Letters (Field Safety Notices) were hand-delivered to U.S. customers on August 19, 2019 and OUS customers on August 20, 2019. Customers were asked to locate affected camera(s) in their inventory and inspect per the instructions attached to the Field Safety Notice throughout the life of the camera(s). If residual soil is present, customers were to return the affected camera(s) to the manufacturer using the standard Return Materials Authorization (RMA) process. Additionally, customers were requested to: iInform necessary hospital personnel about this issue; complete the attached Acknowledgement Form and return it via fax to Intuitive as instructed on the form; and to retain a copy of this letter and the Acknowledgement Form for their files. Customers were informed that an Intuitive Representative would schedule a site visit to review the attached inspection instructions, and that the firm will provide a replacement camera, at no charge, for returned affected product.
Customers requiring further information or support may contact their Clinical Sales Representative or contact Intuitive Customer Service at the following numbers: US Customer Service (800) 876-1310, Option 3 (7 AM to 5 PM EST) or [email protected], or South Korea Customer Service: + 82-02-3271-3200, Option 3 ( 9AM to 6 PM KSTJ)
Root Cause
Process control
Distribution Pattern
Distribution US nationwide and South Korea.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.