GE Pristina Serena - Product Usage: Pristina Serena is an optional accessory of Senographe Pristina intended to provide accurate location of lesions in the breast in three dimensions.

FDA Recall Z-2132-2019 · Posted None

Recalling FirmGE Healthcare, LLC
Recall StatusTerminated
Product CodeMUE
Date Initiated2019-05-16
Date Terminated2020-12-16
Product Quantity16

Reason for Recall

Potential slippage of the biopsy positioner when the gantry is rotated during a biopsy procedure.

Firm Action

On May 17, 2019, GE Healthcare sent letters to all of their consignees advising them of the recall, providing them with new instructions for use, and requesting that they complete the enclosed MEDICAL DEVICE NOTIFICATION ACKNOWLEDGEMENT RESPONSE REQUIRED forms. GE will follow-up and correct these systems,

Root Cause

Equipment maintenance

Distribution Pattern

Worldwide distribution - US nationwide distribution in the states of AZ, CA, FL, GA, IL, MI, MN, NC, NJ, NY, PA, TX, and VT, and countries of Algeria, Australia, Belgium, Canada, Chile, Columbia, Czech Republic, Finland, France, Germany, Hungary, Iceland, Japan, Korea (Republic of), Kuwait, Lebanon, New Zealand, Norway, Poland, Saudi Arabia, Senegal, South Africa, Sri Lanka, Sweden, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, and Vietnam.