JOURNEY II UNI RESECTION PREP TRIAL, LEFT MEDIAL/ RIGHT LATERAL - Product Usage: The JOURNEY II Uni Resection Prep Trial is a reusable surgical instrument used during range of motion trialing to assess joint tension and component alignment during surgery. The provisional moveable spikes (spike rail) aids in provisional fixation of the trial. SIZE 1, CATALOG NUMBER: REF 74035821 SIZE 2, CATALOG NUMBER: REF 74035822 SIZE 3, CATALOG NUMBER: REF 74035823 SIZE 4, CATALOG NUMBER: REF 74035824 SIZE 5, CATALOG NUMBER: REF 74035825 SIZE 6, CATALOG NUMBER: REF 74035826 SIZE 7, CATALOG NUMBER: REF 74035827 SIZE 8, CATALOG NUMBER: REF 74035828 SIZE 9, CATALOG NUMBER: REF 74035829 SIZE 10, CATALOG NUMBER: REF 74035830
FDA Recall Z-2132-2020 · Posted None
Advertisement
Recalling FirmSmith & Nephew, Inc.
Recall StatusTerminated
Product CodeHSX
Date Initiated2020-01-28
Date Terminated2023-12-11
Product Quantity200 units
Reason for Recall
The spike rail was manufactured with an oversized thru-hole which could potentially cause the dowel pin to fall out and the device to disassemble.
Firm Action
The firm initiated the recall by email on 01/28/2020. The initial consignee notified the district sales representatives on 01/28/2020 by email. The district sales representatives checked their inventory and the inventory of the hospital they service.
Root Cause
Process design
Distribution Pattern
US Nationwide distribution.