Dimension Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids.
FDA Recall Z-2133-2014 · Posted 2014-07-30
Advertisement
Recalling FirmSiemens Healthcare Diagnostics, Inc.
Recall StatusTerminated
Product CodeJJE
Date Initiated2014-05-27
Date Terminated2014-12-15
Product Quantity15
Reason for Recall
There is a misassembly issue with the
photometer installed on Dimension Integrated Chemistry Systems instruments. Through internal investigation by
Siemens Healthcare Diagnostics has confirmed that a limited number of Dimension photometers were misassembled. These photometers are performing within specifications, but may lead to a shortened life of the source lamp.
Firm Action
An urgent field safety notice, dated May 2014, was sent to users to inform them of the issue and that a service representative would replace the misassembled photometer.
Root Cause
Employee error
Distribution Pattern
FL LA AR IA OH OH WI IL LA KS MT
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
K130276DIMENSION EXL WITH LM SYSTEMSiemens Healthcare Diagnostics, Inc. · 2013-03-22K112999DIMENSION CLINICAL CHEMISTRY SYSTEMSiemens Healthcare Diagnostics · 2011-11-22K073604DIMENSION EXL WITH LM CLINICAL CHEMISTRY SYSTEM, DIMENSION FT4L FLEX REAGENT CARTRIDGE, AND LOCI THYROID CALIBRATOR, MODDade Behring, Inc. · 2008-03-13