Dimension Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids.

FDA Recall Z-2133-2014 · Posted 2014-07-30

Recall StatusTerminated
Product CodeJJE
Date Initiated2014-05-27
Date Terminated2014-12-15
Product Quantity15

Reason for Recall

There is a misassembly issue with the photometer installed on Dimension Integrated Chemistry Systems instruments. Through internal investigation by Siemens Healthcare Diagnostics has confirmed that a limited number of Dimension photometers were misassembled. These photometers are performing within specifications, but may lead to a shortened life of the source lamp.

Firm Action

An urgent field safety notice, dated May 2014, was sent to users to inform them of the issue and that a service representative would replace the misassembled photometer.

Root Cause

Employee error

Distribution Pattern

FL LA AR IA OH OH WI IL LA KS MT