Terminated

JOURNEY II UNI RESECTION PREP TRIAL, RIGHT MEDIAL/ LEFT LATERAL - Product Usage: The JOURNEY II Uni Resection Prep Trial is a reusable surgical instrument used during range of motion trialing to assess joint tension and component alignment during surgery. The provisional moveable spikes (spike rail) aids in provisional fixation of the trial. SIZE 1, CATALOG NUMBER: REF 74035831 SIZE 2, CATALOG NUMBER: REF 74035832 SIZE 3, CATALOG NUMBER: REF 74035833 SIZE 4, CATALOG NUMBER: REF 74035834 SIZE 5, CATALOG NUMBER: REF 74035835 SIZE 6, CATALOG NUMBER: REF 74035836 SIZE 7, CATALOG NUMBER: REF 74035837 SIZE 8, CATALOG NUMBER: REF 74035838 SIZE 9, CATALOG NUMBER: REF 74035839 SIZE 10, CATALOG NUMBER: REF 74035840

FDA Recall Z-2133-2020 · Posted None

Recalling FirmSmith & Nephew, Inc.
Recall StatusTerminated
Product CodeHSX
Date Initiated2020-01-28
Date Terminated2023-12-11
Product Quantity195 units

Reason for Recall

The spike rail was manufactured with an oversized thru-hole which could potentially cause the dowel pin to fall out and the device to disassemble.

Firm Action

The firm initiated the recall by email on 01/28/2020. The initial consignee notified the district sales representatives on 01/28/2020 by email. The district sales representatives checked their inventory and the inventory of the hospital they service.

Root Cause

Process design

Distribution Pattern

US Nationwide distribution.