Terminated

Drystar AXYS, Hardcopy Printer A free standing device used to print diagnostic conventional and mammography images on transparent film for viewing on a standard view box. It may be used in any situation in which a hard copy of an image generated by a medical imaging device is required or desirable.

FDA Recall Z-2136-2008 · Posted 2008-09-18

Recalling FirmAGFA Corp.
Recall StatusTerminated
Product CodeLMC
Date Initiated2008-05-16
Date Terminated2008-10-10
Product Quantity33 units

Reason for Recall

The "Film Calibration" setting on the printers was set to the default "OFF" position instead of "ON".

Firm Action

Consignees were notified by an Urgent Safety Notice letter on/about 05/16/2008. The letter provides the procedure the customer can follow to change the film calibration setting from "OFF" to "ON". The letter also provides information on how to get assistance from Agfa, if the customers prefer a service engineer perform the change. Consignees were instructed to send a copy of the letter or similar letter to their affected accounts. Agfa also offers to dispatch a service engineer to perform the parameter change.

Root Cause

Other

Distribution Pattern

Worldwide Distribution --- USA including states of DE, FL, IL, KS, MA, MD, ME, MO, MT, NJ, NY, OH, OK, TX, UT, VA, WA, WV, and Puerto Rico and country of Canada.