cobas e 801 immunoassay analyzer, cobas 8000 Modular Series The device is a fully automated, software controlled analyzer system for in vitro determination of analytes in human body fluids. It is part of the cobas 8000 modular analyzer series. It uses electrochemiluminescent technology for signal generation and measurement.
FDA Recall Z-2136-2018 · Posted None
Advertisement
Recalling FirmRoche Diagnostics Corporation
Recall StatusTerminated
Product CodeJJE
Date Initiated2018-04-03
Date Terminated2019-11-21
Product Quantity35
Reason for Recall
The syringe plunger on the device may be mounted in a tilted position, which applies mechanical stress to the plunger. In the worst case scenario, the plunger may crack leading to impaired syringe function.
Firm Action
On April 3, 2018, The Firm sent an Urgent Medical Device Correction letter via UPS Ground.
Consignees were notified of the possible failure. Consignees were informed that a Roche representative would contact them to install a syringe plunger guide for the affected syringes to resolve the issue. In the meantime, consignees were instructed to only measure assays that do not require prewash steps and assays that are not impacted by missing prewash step until the syringe plunger guide is installed on the affected module. If the customer identifies broken syringe plungers on their device, the notice instructed them to contact the Roche Support Network Customer Support Center at 1-800-428-2336, available 24 hours a day, 7 days a week. Customers were asked to complete a customer response form and return it to the firm.
Root Cause
Process design
Distribution Pattern
US Distribution to the states of : CA, CT, IA, IN, KY, ME, MI, NC, NY, OH, SC, TX, and VA.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement