Terminated

AdvanDx Quick FISH Fixation Kit for BC ( Blood Cultures), 50 tests. In vitro diagnostic. The QuickFISH Fixation Kits are components that are sold separately for the Staphylococcus QuickFISH BC test Catalog number: QFFixBC1-50 (a component for QFSTABC1-50)

FDA Recall Z-2138-2014 · Posted 2014-08-01

Recalling FirmAdvandx Inc
Recall StatusTerminated
Product CodeNXX
Date Initiated2014-07-04
Date Terminated2014-09-30
Product Quantity10 kits

Reason for Recall

Positive result in the Internal negative control well for the Staphylococcus QuickFISH BC test, failing QC

Firm Action

AdvanDx notified customers by e-mail on July 4, 2014 of the problem and advised to discontinue use of the QuickFISH Fixation Kits with Lot#04914. Replacement kits were immediately shipped to all of the affected customers. Questions contact local AdvanDx representative or Technical Support 1-866-376-0009, fax 1-781-376-0111 or e-mail: [email protected]. An acknowledgement of Receipt and Destruction of Materials requested to be completed.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Distributed in the states of AZ, ID, FL, MO, OH, MI, NJ, and VA.