RayStation (Radiation Treatment Planning System) : RayStation 9A, 9B, 9B SP1, 10A, 10A SP1, 10B and 11A, in combination with RayCare Model Numbers/ UDI: RayStation 9.0 (9.0.0.113) 07350002010174; RayStation 9.1 (9.1.0.933) 07350002010266; RayStation 9.1 Service Pack 1 (9.2.0.483) 07350002010297; RayStation 10.0 (10.0.0.1154) 07350002010303; RayStation 10.0 Service Pack 1 (10.0.1.52) 07350002010365; RayStation 10.1 (10.1.0.613) 07350002010310; RayStation 11.0 (11.0.0.951) 07350002010389;
FDA Recall Z-2138-2021 · Posted None
Advertisement
Recalling FirmRAYSEARCH LABORATORIES AB
Recall StatusTerminated
Product CodeMUJ
Date Initiated2021-06-08
Date Terminated2024-03-20
Product Quantity5 units
Reason for Recall
Initial delivery positions will be set incorrectly when having setup beam(s) at the localization point.
Firm Action
RaySearch issued Urgent Field Safety Notice, Medical Device Correction #80230 via email on 06/08/21. Letter states reason for recall, health risk and action to take:
TO BE TAKEN BY THE USER
" Do not create setup beams with the localization point as isocenter (option 2).
" Always perform secondary verification of new treatment positions. RayTreat will notify the user when the treatment positions have been updated.
" Always perform imaging before treating the patient.
" Educate planning staff and all users about this workaround.
" Inspect your product and identify all installed units with the above software version number(s).
" Confirm you have read and understood this notice by replying to the notification email.
SOLUTION
This issue will be resolved in the next version of RayStation, scheduled for market release in June 2021 (subject to market clearance in some markets). If customers wish to continue using versions of RayStation affected by this notice, all users must maintain awareness of this notice. Alternatively, customers can choose to upgrade to the new version once it becomes available for clinical use.
For regulatory information, please contact [email protected].
Root Cause
Under Investigation by firm
Distribution Pattern
US distribution to TN; and Belgium
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.