Heater Cooler Unit (HCU 30) The device supplies temperature-controlled water for cardioplegia heat exchangers and for blood heat exchangers in extracorporeal circulation and for blankets with which patients can be warmed or cooled.

FDA Recall Z-2140-2021 · Posted None

Recall StatusTerminated
Product CodeDWC
Date Initiated2021-07-14
Date Terminated2022-08-03
Product Quantity33

Reason for Recall

There is the potential for contaminated water to enter parts of the Heater-Cooler Unit circuits. There is a potential risk of bacterial contamination including Mycobacteria species.

Firm Action

On July 14, 2021, Maquet Getinge Group issued Urgent Medical Device Recall notices to customers via FedEx 2 day Delivery with Signature Proof of Delivery. Customers were advised to take the following action: - Users should take the unit out of operation at the earliest opportunity and decommission the unit. - MCP has not developed a cleaning protocol that meets current concerns and expectations of the FDA. - If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. - Please acknowledge receipt of the Urgent Medical Device Recall by completing and returning the response form. Please either fax the completed form to (866) 5715830 or send via email to [email protected].

Root Cause

Device Design

Distribution Pattern

Distribution to US states of AL, CA, FL, IL, KS, MI, MO, NC, NE & OH.