Accelerator a3600 Automation System with Advia2120LAS Interface Module (P/N FLX-219-00, FLX-219-10), ImmunoCAP 1000 Interface Module (P/N FLX-226-01, FLX-226-10), or StaRRsed Interface Module (P/N FLX-268-00) is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.
FDA Recall Z-2141-2020 · Posted None
Advertisement
Recalling FirmInpeco S.A.
Recall StatusTerminated
Product CodeJQP
Date Initiated2020-04-07
Date Terminated2024-02-06
Product Quantity3 systems
Reason for Recall
The identified problem is an erroneous association between the carrier and the sample tube caused by a communication error between the firmware of the Interface Modules and the Automation software. This problem can occur only when one of these Interface Modules is put off-line after a carrier in their secondary lane is physically returned on the main track and then put back on-line when the carrier is used to transport another tube.
Firm Action
Urgent: Medical Device Correction notification letters dated 3/30/20 were sent to customers.
1) Visually check that the secondary lane of the Interface Modules listed in Table 1 is empty before sending the Off-line command; or
2) Select the "Going to Off-line" command for the Interface Modules listed in Table 1. This ensures that the
Module completes processing samples already inside the Module, releases the tubes and then passes to Off-line status.
Your service provider will contact you to schedule the firmware upgrade.
Until the service visit please maintain awareness on this notice.
Please complete and return the "Customer Letter Receipt Confirmation and Implementation Check form" attached to this letter within 15 days directly to the email address specified in the email communication.
Please transfer this notice to whom it might concern.
Contact reference person:
For any clarification you may need, do not hesitate to contact:
Eva Balzarotti - Regulatory Affairs Manager
E-mail: [email protected]
Phone: (+41) 919118 224
Root Cause
Software design
Distribution Pattern
US Nationwide including in the states of IL, NY.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.