CERTAIN(R) BELLATEK(R) TITANIUM ABUTMENT 3.4MM, Item Number IIAT3 - Product Usage: BIOMET3i Restorative products are intended for use as an accessory to endosseous dental implants for placement in the maxilla and mandible
FDA Recall Z-2142-2019 · Posted None
Advertisement
Recalling FirmBiomet 3i, LLC
Recall StatusTerminated
Product CodeNHA
Date Initiated2019-02-26
Date Terminated2020-06-15
Product Quantity6 units
Reason for Recall
During manufacturing, the screw stop ledge was not made to specification and location within the abutment, resulting in the screw head seating deeper in the abutment with no positive stop.
Firm Action
The firm disseminated the notices by email on 02/26/2019. The letter requested that the dental laboratories quarantine and return any unused product and to further notify the affected clinician referral.
Root Cause
Process control
Distribution Pattern
Nationwide and Puerto Rico, Canada, and Australia
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.