Terminated

Merge OfficePACS software. The firm name on the label is Merge Healthcare, Hartland, WI. OfficePACS is intended for the manipulation and displaying of medical images by a suitable licensed and qualified healthcare professional. It can show images from different modalities and interfaces to various image storage and printing devices using DICOM or similar interface standards. OrthoPACS also uses such images, in conjunction with templates for prosthetic devices, for the purposes of choosing the nature and characteristics of the prosthetic device to be used when planning a potential surgical procedure.

FDA Recall Z-2144-2016 · Posted 2016-07-08

Recall StatusTerminated
Product CodeLLZ
Date Initiated2016-01-30
Date Terminated2017-02-03
Product Quantity216 sites have the affected software

Reason for Recall

Potential data loss occurs as a result of the software archiving not working properly.

Firm Action

Merge Healthcare sent an Urgent Medical Device Recall letter dated January 26, 2016, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customers. Customers were instructed to provide a copy of the notification to all users. For additional questions customers were advised to call Merge Customer Service at (87) 741-5369. For questions regarding this recall call 262-367-0700.

Root Cause

Software design

Distribution Pattern

The software was distributed nationwide to medical facilities. There was no government/military/foreign distribution.