Terminated

GE Stereotaxy Positioner, model 2405544-3, for use with Senographe DS Full Field Mammography system, models 2383168, 2383168-2, 2383168-3, 2383168-3-1, 2383168-4-1. The expected use of the Senographe DS Stereotaxy is an optional accessory for the Senographe system for mammography examinations.

FDA Recall Z-2147-2008 · Posted 2008-09-26

Recalling FirmGE Healthcare
Recall StatusTerminated
Product CodeIZH
Date Initiated2008-07-01
Date Terminated2011-12-11
Product Quantity12

Reason for Recall

GE Healthcare has recently become aware of x-ray emission beyond the edge of the detector primary barrier. This issue occurs when an exam is performed in a specific angulated view associated with use of the Stereotactic Positioner of your Senographe DS Acquisition system and could impact patient safety. If the Stereotactic Positioner is removed when the examination arm is at 33 degrees and, th

Firm Action

A Product Safety Notification letter, dated March 3, 2008, was sent to affected customers. The letter provided information about the safety issue, identified affected product and stated that GE Healthcare will provide a software upgrade. Contact GE Healthcare at 1-262-544-3894 for assistance.

Root Cause

Radiation Control for Health and Safety Act

Distribution Pattern

AL, CA, CO, FL, IL, KS, MA, ME, MN,MO, NC, NJ, NM, NY OK, SC, TN, TX, and WI.