Terminated

Maquet Cardiopulmonary (MCP)/Getinge HLS Set Advanced 7.0. Model Number BEQ-HLS-7050 USA, Part Number 701052794

FDA Recall Z-2149-2019 · Posted None

Recall StatusTerminated
Product CodeDTZ
Date Initiated2019-07-11
Date Terminated2020-12-11
Product Quantity54

Reason for Recall

The sets are configured with quick connectors that have been assembled in reverse on the arterial and venous lines, resulting in blue (venous) to red (arterial) and red (arterial) to blue (venous) connections. Due to this incorrect assembly, the sets cannot be primed from the reservoir prior to use, and the device may not function as intended.

Firm Action

Urgent Medical Device Recall - Removal notification letters dated 7/11/19 were sent to customers.

Root Cause

Under Investigation by firm

Distribution Pattern

The products were distributed to the following US states: AR, CA, FL, GA, IA, IL, IN, KY, NC, NE, NM, NV, NY, OH, OR, TN, and WI.