Terminated

Human Assayed Multi-Sera Level 2, Model NO. HN1530 Product Usage: This product is intended for in vitro diagnostic use, in the quality control of diagnostic assays. The Human Assayed Multi-sera is for the control of accuracy.

FDA Recall Z-2150-2019 · Posted None

Recall StatusTerminated
Product CodeJJY
Date Initiated2019-06-11
Date Terminated2020-05-15
Product Quantity149 (US)

Reason for Recall

The RX Series control target and range value for ALT (Tris buffer without P5P 37C), Bicarbonate and Gamma-GT has been incorrectly assigned in specific lots of the Human Assayed Multi-Sera Controls.

Firm Action

On May 25, 2019, the firm notified its customers of the recall via Urgent Medical Device Correction letters. Customers were instructed to do the following: " Review your reagent inventory of these products. Remove all previous versions of the Value Sheets in use and replace with the revised version. " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to t echnical.services@ randox.com within five working days. If you have any questions or concerns, please contact Randox Technical Services.

Root Cause

Under Investigation by firm

Distribution Pattern

US Nationwide Distribution - AZ CA CT FL GA IL KY LA MA MD MI MN MO NJ NY OH OK OR PA TX UT VA WA WI WV and PR. Worldwide foreign distribution.