PopLok Punch labeled in part: "PKL-35M***3.5mm***QTY 1***PopLok Punch***ConMed Linvatec Biomaterials, Ltd. Hermiankatu 6-8L 33720 Tampere, FINLAND***Rx Only***ConMed Linvatec 11311 Concept Blvd Largo, FL 33773-4908 USA***Made in USA***" Intended Use: Bone Punch.
FDA Recall Z-2151-2012 · Posted 2012-08-06
Advertisement
Recalling FirmLinvatec Corp. dba ConMed Linvatec
Recall StatusTerminated
Product CodeLXH
Date Initiated2012-07-06
Date Terminated2013-11-21
Product Quantity2
Reason for Recall
ConMed Linvatec recalled one lot of their PopLok bone Punch 3.5MM because the metal shaft diameter is 4.5mm when it should have been 3.5mm.
Firm Action
The recall is being conducted to the end user level. ConMed Linvatec Urgent Medical Device Recall Notifications were sent by FedEx next day delivery or equivalent method to notify all direct consignees. The notifications identified the affected product along with the reason for the recall and health risk. All accounts were asked to immediately check their facility for the affected product, stop using any of the product with specified lot number, and to segregate the product. The letter also asked customers to return all affected product to ConMed Linvatec and respond using a Reply Form, even if there is no affected product available. If customers have any additional questions, they should contact the Customer Service department at 800-237-0169.
Root Cause
Process change control
Distribution Pattern
Worldwide Distribution -- US and Korea.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.