Pausch Uroview FD fluoroscopic X-ray system UDI number (01)04260491620040 Primarily for urological applications (functional X-ray diagnostics, endourology and minimal invasive urology / surgery). The system, which includes a radiologic / urologic treatment table, maybe used for urological treatment, planning and diagnostic procedures, including but not limited to.
FDA Recall Z-2152-2017 · Posted None
Advertisement
Recalling FirmHans Pausch Rontgengeratebau Gmbh
Recall StatusTerminated
Product CodeJAA
Date Initiated2017-03-14
Date Terminated2019-06-14
Product Quantity10 systems
Reason for Recall
During patient interventions with the urology table the radiation is activated with a footswitch connected to the table. This footswitch is actively involved in the safety circuit of the urological table. Only if footswitch + exposure + first primary barrier are closed the signal will be transmitted to the generator for activating X-ray. During intended use only with this footswitch X-ray should be activated.
Firm Action
Pausch Medical will bring defect into compliance :
The Center for Devices and Radiological Health (CDRH) understands that the CAP involving 10 units will consist of the following:
1. You will contact customers and initiate the hardware upgrade to correct the noncompliance.
2. You will install the hardware upgrade and perform testing to ensure this update was effective.
3. Your customer notification will include a statement that you will without charge, bring the product into compliance.
CDRH approves the CAP subject to the following conditions:
1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification.
2. You will send a list of the product locations and associated telephone number to the FDA district office.
3. You will implement this CAP by July 1, 2017.
You may proceed or continue with implementation of the CAP (subject to the conditions noted above).
All actions to install this retrofit kit 01164901 have been executed until Mar 13,2017. The additional retrofit kit 01164901 was extended to additional batches of the same Uroview FD device.
Root Cause
Device Design
Distribution Pattern
Nationwide Distribution