Recall StatusOpen, Classified
Date Initiated2024-05-23
Product Quantity209 units
Reason for Recall
The design of the HRIS Storage & Sterilization Case and Tray P/N 6210-9-900 and the Gray Revision Instrument System Accessory Tray/Case P/N 6210-9-910 may lead to misuse of the Gray Revision Instruments including re-use and re-sterilization of the single-use, sterile-packaged instruments.
Firm Action
On May 23, 2024 MEDICAL DEVICE RECALL letters were sent to customers. Actions needed: Our records indicate that you may have received the affected product(s). It is Stryker s responsibility as the
manufacturer to ensure that customers who may have received these affected products also receive this
important communication. We therefore request that you read this notice carefully and complete the
following actions.
1. Please inform users of this Medical Device Recall and forward this notice to all individuals who need
to be made aware.
2. Immediately check all stock areas and/or operating room storage at your facility to determine if any
Gray Revision Instruments from the affected product list in Table 2 have been removed from sterile
packaging (Gray Revision Instruments are single-use, sterile-packaged manual surgical instruments).
3. If any Gray Revision Instruments have been removed from sterile packaging, discard/dispose of those
instruments.
4. Gray Revision Instruments found within their sterile packaging may be used as intended (single-use,
sterile-packaged).
5. Do not use the HRIS Storage & Sterilization Case and Tray or the Gray Revision Instrument System
Accessory Tray/Case to re-process (re-sterilize, clean, or re-use) the Gray Revision Instruments,
which are single-use, sterile-packaged manual surgical instruments.
6. You may continue to use the HRIS Storage & Sterilization Case and Tray or the Gray Revision
Instrument System Accessory Tray/Case for all other instruments as intended.
7. Complete and sign the enclosed Medical Device Recall Business Reply Form and email to
[email protected]/ fax (844) 412-9273. 8. Please contact your Local Sales Office or your Stryker Sales Representative directly for
product replacement and inventory questions. Please assist us in meeting our regulatory obligation by emailing back the
attached Medical Device Recall Business Reply Form within 5 days. A response
is required, even though you may not have any physic
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Brazil, Canada, Chile, Denmark, Hong Kong, Iran, Italy, Japan, Kuwait, Mexico, New Zealand, Singapore, Spain, Sweden, Switzerland, Thailand, Turkey, UK, Venlo.