3.5mm X 32.5mm Cortical Screw used with the Polarus Humeral Fixation System,. Product labeled in part, "Polarus Humeral Fixation System& SIZE: 3.5mm X 32.5mm Cortical Screw... REF HCO3325-S... Acumed 5885 NW Cornelius Pass Road Hillsboro, OR 97124-9432 ". Polarus Humeral Fixation System, 3.5mm X 32.5mm Cortical Screw are packaged in a sterile inner package which are placed into an outer box. The cortical screws are used with the Polarus Locking Humeral Rod (Polarus Huemral Fixation System) for the multi-planar fixation. The screws are sold individually to support the Polarus Locking Humeral Rod systems in the field.

FDA Recall Z-2155-2008 · Posted 2008-08-28

Recalling FirmAcumed LLC
Recall StatusTerminated
Product CodeHWC
Date Initiated2005-12-27
Date Terminated2010-02-03
Product Quantity112 pieces

Reason for Recall

Package of 3.5mm X 32.5mm cortical screw may contain 3.5mm X 22.5mm cortical screw.

Firm Action

The firm issued a Product Removal notification on 12-27-2005 to their customers. The notification was sent by either FAX or by e-mail. The Product Removal Notification informed their customers of the issue, asked them to remove the screws from inventory and return the screws to Acumed. A reminder notification, dated 1/20/08 was issued to customers identifying the product and the problem and requesting the product be returned.

Root Cause

Process change control

Distribution Pattern

Worldwide Distribution --- USA including states of CA, KY, FL, TX, OH, OR, NY, ID NJ, NE, LA, WI, UT, MO, MS, GA, AZ and countries of United Kingdom, Italy, Puerto Rico, Turkey, Spain, Japan, South Africa, and Belgium.