CytoGuard Closed Luer Connector, Intravenous access 2 cartons of 50 units each (100 units per case)
FDA Recall Z-2155-2017 · Posted None
Advertisement
Recalling FirmB Braun Medical Inc
Recall StatusTerminated
Product CodeFPA
Date Initiated2017-04-10
Date Terminated2018-02-02
Product Quantity417,600 units
Reason for Recall
B. Braun is voluntarily recalling 13 lots of CytoGuard Closed Luer Connector due to a potential for some blisters to be punctured resulting in a compromised sterility barrier.
Firm Action
B. Braun mailed an Urgent Medical Device Recall Notice to affected customers to inform them of the issue. Customers were asked to determine their current inventory and to not destroy any of the product; however, product should be discontinued immediately and quarantined and return the acknowledgement form for product return.
Root Cause
Nonconforming Material/Component
Distribution Pattern
United States Nationwide distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.