Sonopet Ultrasonic Surgical System Console, Ultrasonic Aspirator, Stryker Medtech.

FDA Recall Z-2156-2014 · Posted 2014-08-07

Recall StatusTerminated
Product CodeLFL
Date Initiated2014-06-24
Date Terminated2015-10-27
Product Quantity2,460 units

Reason for Recall

The firm received complaints regarding insufficient irrigation to the tip of the Sonopet system. Lack of irrigation can potentially lead to excessive heat at the tip.

Firm Action

The firm sent Medical Device Upgrade Notification letters dated July 11, 2014 to their customers.

Root Cause

Labeling design

Distribution Pattern

Nationwide Distribution-including DC and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI & WV.