Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Medline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) NERVE BLOCK ACCESSORY PACK, Model Number DYNJRA1220; b) PERI NEURO BLOCK PACK-LF, Model Number DYNJRA1773; c) PERI NEURO BLOCK PACK-LF, Model Number DYNJ51107B; d) PERI NEURO BLOCK PACK-LF, Model Number DYNJ0703127A; e) PERI NEURO BLOCK PACK-LF, Model Number DYNJ42367B; f) PERI NEURO BLOCK PACK-LF, Model Number DYNJRA1571; g) CVOR ANGIOGRAPHY PACK-LF, Model Number DYNJRA0656A; h) CVOR ANGIOGRAPHY PACK-LF, Model Number DYNJ47338; i) BLOCK KIT-LF, Model Number DYNJRA1287A; j) BLOCK KIT-LF, Model Number PAIN1622A; k) BLOCK KIT-LF, Model Number DYNJRA1543; l) PK, RADIOLOGY-NEURO-IR, Model Number DYNJRA1635A; m) PK, RADIOLOGY-NEURO-IR, Model Number DYNJ0134147D; n) PK, RADIOLOGY-NEURO-IR, Model Number DYNJRA1604; o) BLOCK TRAY, Model Number PAIN1182;

FDA Recall Z-2158-2023 · Posted 2023-07-20

Recall StatusOpen, Classified
Product CodeOGJ
Date Initiated2023-05-18
Product Quantity23938 units

Reason for Recall

Probe covers for use during diagnostic ultrasound procedures may have an inadequate barrier at the seams.

Firm Action

Medline distributed their recall notices to their consignees beginning on 05/18/2023 by letter. The notices explained the problem, and the risk and requested the destruction of the products. Distributors were directed to notify their consignees, have the customers document and destroy any affected product, and include your customers quantities on the response form. Beginning on 05/23/2023, Medline issued another notice to a separate group of consignees requesting that the kits be over labeled. Distributors were directed to notify the customers and have them return any affected product to the distributor for collection and return to Medline. Questions: please contact the Recall Department at 866-359-1704

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution: US (Nationwide) and OUS (foreign) United Arab Emirates, Singapore, Lebanon, China Qatar, Costa Rica, India, Canada, Panama & Saudi Arabia.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑