Inspire Model 3028, IV Implantable Pulse Generator
FDA Recall Z-2159-2024 · Posted 2024-06-21
Advertisement
Recalling FirmInspire Medical Systems Inc.
Recall StatusOpen, Classified
Product CodeMNQ
Date Initiated2024-05-16
Product Quantity24 devices
Reason for Recall
There is a potential manufacturing defect which can lead to electrical leakage in the sensing circuit which causes a system malfunction and requires a revision surgery to replace the IPG,
Firm Action
Inspire Medical systems issued an URGENT Medical Device Correction notice to the affected healthcare providers on 05/16/2024 via FedEx overnight. The notice explains the potential problem with the device and risk to the patient and recommends the following actions:
We recommend the following actions:
" Notify affected patients of this voluntary recall.
" Schedule your patient for a monitoring visit to confirm their Inspire therapy is working as intended through waveform and impedance analysis. Monitor patients for reports of variation in stimulation, a lack of therapy effectiveness, or an inability to turn on therapy.
" Continue with routine monitoring and conduct waveform and impedance analysis at every routine office visit, as IPG issue may be non-invasively identified with basic waveform and impedance testing.
" If abnormal waveforms or impedances are observed, record impedance values at 1.5V, 2.0V, 2.5V, 3.0V and 3.5V.
o Let the patient know they will feel 3 short pulses of stimulation.
o Some of these stimulation pulses may feel strong. This is normal.
" Contact your Inspire representative for additional diagnostic testing and troubleshooting recommendations.
For additional questions: please contact your Inspire representative or the Inspire Quality team at [email protected].
Root Cause
Process control
Distribution Pattern
Worldwide distribution - US Nationwide and the country of Germany.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.