Hoffmann LRF Safety Clip (Bone Transport Strut), 14 x 12 x 5 mm, REF 4933-0-401, NDC (01)07613327131475(10)V07241 Product Usage: Intended for fixation of fractures, joint contractures, fusions, limb lengthening, deformity correction, and bone and soft tissue reconstruction in pediatric patients and adults.

FDA Recall Z-2160-2017 · Posted None

Recall StatusTerminated
Product CodeKTT
Date Initiated2017-04-26
Date Terminated2020-02-07
Product Quantity437 units

Reason for Recall

Stryker is recalling Hoffmann LRF (Limb Reconstruction Frame) System Safety Clip - Bone Transport Strut because there is a potential for loosening of the safety clips from the intended position on the Quick Release mechanism of the Transport Struts.

Firm Action

Stryker sent via FedEx an "Urgent Medical Device Recall Notification and Business Reply Form" dated April 26, 2017 to their affected customers. For questions call (201) 831-6693.

Root Cause

Device Design

Distribution Pattern

US Nationwide distribution in the states of CA, FL, MD and VT