BD COR System Software. Model Number: 444829.
FDA Recall Z-2162-2025 · Posted 2025-07-23
Advertisement
Recalling FirmBecton Dickinson & Co.
Recall StatusOpen, Classified
Product CodeMAQ
Date Initiated2025-06-23
Product Quantity7 systems
Reason for Recall
Potential for functionality issue that supports the Over
labeling feature and could lead to possible false negative results.
Firm Action
Becton, Dickinson and Company began notification of consignees on about 06/23/2025 via telephone call and followed by a letter. Consignees were instructed to note the outlined workflows when loading specimens with the Over labeling feature enabled and consider retesting affected specimens for HPV testing if affected. Consignees were also instructed to post the notification within their facility, notify customers if affected units were further distributed, and complete and return the provided Customer Response Form.
Root Cause
Under Investigation by firm
Distribution Pattern
Worldwide - US Nationwide distribution in the states of Florida, Texas and Michigan. The countries of Sweden and Belgium.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.