Diowave Laser System, REF: Diowave 250W
FDA Recall Z-2162-2026 · Posted 2026-05-11
Advertisement
Recalling FirmTechnological Medical Advancements LLC
Recall StatusOpen, Classified
Product CodeILY
Date Initiated2026-01-09
Product Quantity16
Reason for Recall
Software update is needed for laser systems, for topical infrared heating to provide temporary relief of minor muscle and joint pain, spasms, stiffness associated with minor arthritis, and to temporarily increase local blood circulation, operating at a maximum average power output of 70W, which is beyond the FDA cleared 60W.
Firm Action
Starting on 1/9/2026, customer notifications were sent informing customers to install software update V1.81.
On 3/11/2026, correction notices were sent to customers informing them to take the following actions:
1) Install Software Update V1.81. Follow the WiFi upgrade instructions to complete the update. The update is free of charge and requires no hardware modifications.
2) Confirm Completion. Once the update is installed, please complete and return the enclosed Correction Response Form by email to [email protected]. Alternatively, mail the completed form to TMA at the address above.
3) Forward This Notice. This notice must be passed on to all persons within your organization who need to be aware of this correction, and to any organization where the affected device has been transferred or is currently in use.
4) Contact Support If Needed. Firm's technical support team is available Monday through Friday, 9:00 AM 5:00 PM ET, at (561) 722-1153 or [email protected]
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide distribution in the states of TX, CA, MI, MD, PA, FL, MN, DC, MI.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.