SENSATION PLUS 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK Product Code/Part Number: 0684-00-0576-01U is a cardiac assist device which support the heart's left ventricle by increasing coronary perfusion and reducing the work of the left ventricle.
FDA Recall Z-2164-2017 · Posted None
Advertisement
Recalling FirmDatascope Corporation
Recall StatusTerminated
Product CodeDSP
Date Initiated2017-04-12
Date Terminated2020-01-28
Product Quantity996 UNITS
Reason for Recall
Maquet/Getinge is initiating a voluntary product removal involving four lot numbers of SENSATION PLUS 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK since during internal pressure (leak) testing of the SENSATION PLUS 8Fr. 50cc, IABC, failures related to the tip seal leak were identified.
Firm Action
Getinge notified their customers with a recall letter/return response form via Fed Ex on 4/12/2017.
Getinge notified their customers again through an expansion on 10/16/2017.
Root Cause
Device Design
Distribution Pattern
Nationwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.